Document Type

Conference Proceeding

Publication Date

2-5-2002

Publisher

American Institute of Physics

Abstract

The Cryogenic Dark Matter Search (CDMS) employs detectors which are capable of simultaneously measuring the ionization and phonon energies deposited by a particle collision. These detectors are 1-cm-thick, 7-cm-diameter crystals of either germanium or silicon with a thin film of aluminum and tungsten patterned on the surface. This presentation discusses the testing regimen that a typical CDMS detector undergoes before it gets approval for final installation at the CDMS II deep site in Soudan, MN which will come online in early 2002. Now that our technology is relatively stable, the main focus of our test facilities is to provide quality control for the mass production of our detectors. First, the critical temperatures of the tungsten and other basic quantities are measured in preparation for iron implantation, which will bring the Tc down to the desired range ( 70 mK). The same basic measurements are taken again after implantation to assure that the correct Tc was achieved. Finally, a detailed map of energy response as a function of position is made to calibrate residual inhomogeneities across the

Editor

F. Scott Porter
Dan McCammon
Massimiliano Galeazzi
Caroline K. Stahle

Comments

Copyright © 2002 American Institute of Physics Publishing. Reprinted with permission.

Low Temperature Detectors: Ninth International Workshop on Low Temperature Detectors (Madison, Wisconsin, July 2001)

Included in

Physics Commons

Share

COinS
 
 

To view the content in your browser, please download Adobe Reader or, alternately,
you may Download the file to your hard drive.

NOTE: The latest versions of Adobe Reader do not support viewing PDF files within Firefox on Mac OS and if you are using a modern (Intel) Mac, there is no official plugin for viewing PDF files within the browser window.